Jak správně skladovat a používat pH pufrovací roztoky?
After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 stupňů. Obecně lze roztoky pH 7,00, pH 6,86 a pH 14,00 používat po delší dobu. Roztoky s pH 9,18 a pH 10,01 jsou náchylné ke změnám pH v důsledku absorpce oxidu uhličitého ze vzduchu.
Jaký je účel pH pufrovacího roztoku?
Před měřením pH zkalibrujte pH metr.
Pro ověření přesnosti pH metru, například po kalibraci pH metru s pH 6,86 a pH 14,00, vložte pH elektrodu do roztoku pH 9,18 a zkontrolujte, zda hodnota na displeji přístroje odpovídá hodnotě pH standardního roztoku.
Je nutné znovu kalibrovat pH metr při obecném přesném měření. Po kalibraci a použití pH metru může dojít k posunu nebo změnám. Proto před testováním vložte elektrodu do standardního tlumivého roztoku, který je blízko měřenému roztoku, a na základě velikosti chyby určete, zda je nutná rekalibrace.